Customization: | Available |
---|---|
Warranty: | 1 Years |
Type: | Vulcanization Meter Measurement Controller |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
The measurement and control system of the rotorless vulcanizer is specially developed for the determination of rubber vulcanization characteristics. The system adopts advanced PID temperature control technology and switching power supply technology, with accurate measurement, high efficiency and reliable operation. The analysis data of rubber vulcanization characteristics of this measurement and control system can be used as the basis for the quality control of the factory's internal products, and can also be used as a reference for the control of rubber ingredients.
1. Temperature control range: room temperature to 200°C;
2. Temperature resolution: 0.01°C;
3. Temperature control accuracy: ±0.1°C;
4. Temperature recovery speed: <40s;
5. Heating rate: about 35°C/min;
6. Temperature unit: ºC, ºF;
7. Torque range: 0 N·m ~ 20 N·m;
8. Minimum torque reading: 0.01 N m
9. Torque unit: N m, N in, kg m, kg cm, kg in, lb m, lb in, g m, g in, dN m, dN in;
10. Swing angle: suitable for ±0.5°, ±1°, ±3°;
11. Oscillation frequency: suitable for 100r/min (1.66Hz);
12. Power supply: 220V AC;
13. Air source pressure: not less than 0.44MPa (generally 0.3MPa);
14. Hardware interface: USB;
15. Interface language type: Simplified Chinese, Traditional Chinese, English;
16. Quality standard: ISO 6502, GB/T 16584;
17. Report format: Excel, Word, PDF.
System Features
1. The software supports Windows XP, Windows 7, Windows 10 and other operating systems. The software has complete functions (test results can be stored and compared, enlarged, exported, and printed in color, etc.), and can be upgraded according to needs.
2. Years of experience in temperature control and advanced PID self-tuning temperature control technology ensure accurate and stable temperature control.
3. The temperature does not need to be calibrated (it has been calibrated at the factory), and the torque does not need to be calibrated (it has been calibrated at the factory), only the proportional parameter needs to be modified.
Specification :
Dimensions: 215×132×45 (mm) (length×width×depth) Hole size: 158×158 (mm) (length×width) |